• UVT-S-AR PCR Workstation

    The Grant bio UVT-S-AR is a large capacity stainless steel PCR UV cabinet workstation with additional space for equipment and accessories to allow for more comfortable and convenient working in PCR applications.

    These UVT-S-AR cabinets are equipped with an open UV lamp installed in the upper part of the hood. UV-radiation from these lamps effectively disinfects the working area chamber, inactivating DNA / RNA fragments during 15-30 minutes of exposure. The digital timer controls the duration of the direct UV irradiation. An additional daylight lamp provides a bright, clear working area.

    They are equipped with a high efficiency flow-type bactericidal UV cleaner-recirculator which provide constant decontamination inside the cabinet during operation. The UV cleaner-recirculator consists of UV lamp, dust filters and a fan. The recirculation effect increases the maximum density of the UV light making it sufficiently effective for DNA / RNA inactivation. The recirculator processes 360 UV-cabinet volumes per hour, creating permanently aseptic conditions within the cabinet.

    • Robust construction with large working area, 1.2m x 0.52m working area
    • UV surface irradiation – dual 30W 254 nm UV lamp
    • UV protection – UV-protective film on glass panels
    • UV exposure controlled by 24h digital timer
    • Convenient, easy to use digital timer for accurate control of UV exposure
    • Front opening with three adjustable positions for ease of access
    • White lamp provides local illumination of the workplace to optimise visual control during operations
    • Built-in UV cleaner-recirculator both increases the maximum density of UV light and generates 25 m³/h air flow exchange – prevents unwanted contamination and protects the user from direct UV light during manipulation
    • Automatic switch-off of irradiated UV light when door opened
    • Ample additional space for equipment and comfortable working.

    Applications

    These PCR UV-cabinets are designed for clean operations with DNA samples and provide protection against contamination. They are not recommended for operations with dangerous infectious and viral materials as well as DNA / RNA amplification.

    Specifications

    Dimensions (h x d x w) 585 x 585 x 1245mm
    Construction Stainless steel frame and working area
    Panels Glass with UV-protective film
    Front opening with three adjustable positions
    Open UV lamp, 25W bactericidal, 254nm, ozone free
    Open UV lamp, 30W bactericidal, ozone free 2
    Bactericidal air recirculator, 25m3/h air flow exchange
    UV recirculator, 25W (efficiency >99% per 1 cycle)
    UV recirculaor, 30W (efficiency >99% per 1 cycle) 1
    White lamp for workplace illumination – 15W
    White lamp for workplace illumination – 30W 1
    Radiation type Ultraviolet (253.7m), ozone free
    Optical transmission 95%
    Digital timer (non stop) 0 to 24 hours
    Internal power outlets 3
    Internal working area 1200 x 520mm
    Flow rate 70m3/h
    Power consumption – 230V 530W (4.5A)
    Power consumption – 120V 315W (1.4A)
    Nominal operating voltage 120 (60Hz) or 230 (50Hz)
    Weight 58kg