T100 Series Heated Circulating Baths

Grant Optima™ Heating Circulator
= standardGeneral purpose DigitalAdvanced digital
T100TC120TX150TXF200
Stability (DIN 12876) @ 70°C °C 0.05 0.05 0.01 0.01
Uniformity (DIN 12876) @ 70°C °C 0.1 0.1 0.05 0.05
Setting resolution °C 0.1 0.1 0.1 (0.01 with Labwise™)
Display   4 digit LED full colour QVGA TFT
Timer function   - 1 min to 99 hrs 59 mins
No. preset temperatures   3 3 3 3
Re-calibration points   2 2 5 5
Socket for external probe (TXPEP, TXSEP)   - -
Communication interface   - - USB & RS232 USB & RS232
Programmable   - -

remote via PC
1 program/ 30 segments

direct via user interface or remote via PC/laptop
10 programs / 100 segments
Relays   - - 1 1
Safety overtemperature fixed adjustable cut-out
  fluid level - float
Language capability   - - EN, FR, DE, IT, ES EN, FR, DE, IT, ES
Alarms (can be configured to switch a relay)   - high (no relay) high and low high and low
Heater power 230V W 1290 1290 1840 1840
  120V W 1440 1440 1445 1445
Electrical power 230V W 1400 (50-60 Hz) 1400 (50 Hz) 2000 (50 Hz) 2000 (50-60 Hz)
  120V W 1500 (50-60 Hz) 1500 (50-60 Hz) 1500 (50-60 Hz) 1500 (50-60 Hz)
Height above tank rim mm 200 200 200 200
Depth below tank rim mm 135 135 145 145
Grant Optima™ thermostats
Maximum pressure water mbar - 210 310 530
Maximum flow water L/min - 16 18 22 (adjusted flow rate)
Pump connector 6mm bore - fits 9 mm inner diameter tubing
Pump connector 11mm bore - fits 15 mm inner diameter tubing