RTS-1C Personal Bioreactor with Cooling

Personal bioreactors - models and specification
  RTS-1RTS-1C
Theoretically possible measurement range in OD850 , at 10ml working volume*: Rod shaped bacteria 0-25 (0–45.6 OD600 equivalent**)
  Yeast 0-50 (0–75 OD600 equivalent)
E.coli BL21 Factory calibration measurement range, in OD850: at 10–20 ml volume 0 – 10 OD (0 – 19 OD600 equivalent) 
  at 20–30 ml volume 0 – 8 OD (0 – 15.2 OD600 equivalent)
Measurement wavelength (λ)        nm 850 ± 15
Factory calibration measurement precision ±0.3 OD 850
Mass transfer coefficient kLa  (h-1) Up to 350 ± 26 h-1 at 5 ml
Light source   LED
Real time measurement (minutes) 1–60
Temperature setting range °C  +25 to 70 (increment 0.1 °C)  +4  to +70 °C (increment 0.1 °C)
Bottom control range point °C Ambient +5 Ambient -15
Top control range point °C 70  
Stability °C ±0.1
Sample temperature accuracy: 20 °C - 45 °C ± 1
  < 20 °C ± 2
  > 45 °C ± 3
Sample temperature heating/cooling rate °C/min 0.7
Sample volume ml 5 – 30
Speed control range rpm 50 – 2,000 (increment 10 rpm)
Speed control precision rpm ±15
Reverse Spin Time (seconds) 1- 60 (increment 1 s)
Display   LCD
Minimum PC requirements Intel/AMD Processor,
1 GB RAM, Windows XP***/Vista/7/8/8.1/10, 2.0 USB port
Optimal PC requirements Intel/AMD Processor, 3 GB RAM, Windows 7/8/8.1/10,   2.0 USB port
Power consumption 40W (3.3A) 60W (5A)
Input voltage    V dc 12
Nominal operating voltage V 120 - 230 (50/60Hz)
Weight kg 1.7 2.2